Field Sweeping with the SDA-5000: Considerations for Sweep in the Field
79
4
Level Screen
You can set the reference levels from the screen shown in Fig. 4-12, or you
can use the diamond keys.
Call up the level screen from the SWEEP screen (Fig. 4-10) by
pressing the soft key above the Frequency Submenu icon.
Press the soft key next to the Reference Level icon on the right
side of the display for manual reference level adjustment. Use
the up or down diamond keys, or the numeric keypad, to enter
the level, and then press Enter. Note the changes in the graph
as you adjust the values.
Press the Autoscale soft key to auto reference the level. This
automatically sets the scale to a convenient level for optimum
viewing in the center of the screen.
Press the Scale soft key to adjust the graphs scaling. Use the
up and down diamond keys to scroll through the values
available (1, 2, 5, and 10 dB/div). Typically, this value should be
set to 2 dB/div for proper sweep displays. Notice what happens
to the display as you change the scale factor. The change takes
effect from the center of the screen.
Press the Level Submenu icon to return to the main SWEEP
screen.
Fig. 4-12 The SWEEP Level Screen