
Field Sweeping with the SDA-5000: Considerations for Sweep in the Field
71
4
Signal Level Considerations
At amplifier test points, the same basic connection and signal level rules
apply to the SDA-5000 as apply to the headend unit. Too much signal level
on the input of the unit can cause inconsistent sweep readings. The forward
telemetry level should be 0dBmV±12dB at the input of the SDA-5000. This
can be verified by checking the lower right corner of the unit’s display while
sweeping forward (see “Forward Sweep Operation” on page 75). If the
telemetry level is too high, adding in-line attenuators or pads on the unit’s
input is a quick way to reduce signal level.
See Also
½ Chapter 14, “SDA-5500/SDA-5510 Sweep Considerations”
discusses the connection and signal level issues for
headend units.
If both the SDA-5500 and SDA-5510 units are in service, telemetry
frequencies for both the headend and field units must be entered during the
setup of the field units.
NOTE: The field units must also be set to the multiple user
mode (if you’re using the SDA-5510), even if only one of
the maximum ten people are sweeping.
Wiring and Levels: Forward Sweep
Only one cable is required for forward sweep testing. Simply connect a test
point from the output of the amplifier to be aligned or tested to the IN port of
the SDA field unit. If the video carriers in the signal are above +20 dBmV in
the signal after test point loss, then you should use an inline pad to drop the
level to no more than +10dBmV video carrier levels. (Typically this will be the
case only when direct connections to an amplifier output are used.)
NOTE: Remember, as stated previously, too much signal
level on the input can cause inconsistent sweep readings.
The forward telemetry level should be 0dBmV ± 12dB on
the input of the SDA field unit. This can be verified by
checking the lower right corner of the unit’s display while
sweeping forward. Inline pads provide a quick way to
reduce the signal level on the input.
Commentaires sur ces manuels