User Manual 301/302
TRIGger:SOURce:ADVance EXTernal | INTernal
Purpose
This command will set the advance source for the trigger mode.
<EXTernal> will select the external input
<INTernal > will select the internal trigger generator
TRIGger:SOURce:ADVance?
Response
The 302 will return EXT, or INT depending on the present trigger source setting.
TRIGger:TIMer <period>
Purpose
This command will set the period for the internal trigger generator.
<period> numeric value will set the frequency of the internal trigger generator
TRIGger:TIMer?
Response
The 302 will return the present internal period value.
TRIGger:IMMediate
Purpose
This command will stimulate the 302 output once. This command will affect the 302 only after it will be set
to interrupted operation.
*TRG
Purpose
This common command has the same affect as the TRIG:IMM command.