DC AND AC CURRENT MEASUREMENT (See Fig. 3)
Connect red test lead to the A input (ACx for HD115B) for current measurements up to 2A or to the 10A
input for current measurements to 10A. Connect black test lead to COM input connector. Set the Function
Switch to the desired current range. Press the AC/DC menu button to display either AC or DC indicator.
Open circuit in which current is to be measured (voltage between this point and ground must not exceed
1000V). Securely connect test leads in series with the load. Turn on power to circuit being tested. Read
current value on Digital Display.
Note: If, in the 10A range, the meter is exposed to current greater than 10A, it should be turned off and
allowed to cool for at least 10 minutes
Incorrect Input Warning: A tone will sound when a test lead is connected to either amperage jack, but the
selector switch is not set to the correct current range. Ranges 200mA to 2A require the test lead in the A jack
(ACx for HD115B). The 10A range is only for use with the test lead in the 10A jack.
- 14 -
MAX
V
A
10A
COM
CAT III
10A MAX
FUSED
2A MAX
FUSED
1500V
1000V
OFF
AC
DATA
HOLD
DC
V
A
SAFETY
TESTER
TM
1500
1000
200
480V
240V
120V
50V
24V
200m
200m
200u
20m
2m
2A
10A
200
2k
2M
20M
20k
200k
20
2
500V AC
MAX
HD110B
2
0
0
AC
mA
Discharge capacitors
Kondensatoren entladen
Descargue los condensadores
Décharger les condensateurs
4b
4a
4c
4d
2
1
5
6
3
red/rot/
roja/rouge
A or A
Fig. 3
HD110B/15B.Man.09.00 9/11/00 6:00 PM Page 14
Vue de la page 15
1 2 ... 11 12 13 14 15 16 17 18 19 20 21 ... 43 44

Commentaires sur ces manuels

Pas de commentaire